Detector Performance (on the working area)

ParameterValue on first quadrantValue on second quadrant
Read Out Noise (mean)37.4 +/- 0.7 electrons37.7 +/- 0,8 electrons
Dark current (mean)0.347 +/- 0.046 electrons/second0.344 +/- 0/053 electrons/second
Gain4.82 +/- 0.44 electrons/count4.80 +/- 0.45 electrons/count
Bad Pixels2.7 %2.7 %

Last update 25 January 97